ID:20102426169
Category: Manufacturing & Processing Machinery Water treatment & Environment
MOQ:1.00 Piece(s)
Double beam spectrophotometer
T/T
[Features]
◆ Dual Beam Light Path Design
UV-9000 series adopt double light path design, it greatly reduces circuit fluctuation and stray light and ensures high stability of the instrument.
◆ Powerful Stand-alone Operation System
UV-9000 series are euqipped with stand-alone operation system. Functions like spectrum scanning, standard curve, kinetics, multi wavelength scanning, DNA/Protein test, etc. can be directly completed on device without connecting PC software.
◆ Long Light Path Design
The unique 520mm long light path design is adopted to improve the wavelength resolution and make the bandwidth of UV-9000S can be switched to 0.5nm.
◆ UV Analyst PC Software
PC software is provided to connect device with PC for easy data output and analysis, the functions include wavelength scanning, standard curve, kinetics, multi wavelength scanning, DNA/Ptotein test, etc.
◆ 16mm Rigid Aluminim Optical Base
A rigid 16mm thick die-cast aluminim base is used ass the optical mount to ensure high stability and reliability.
◆ One-button Automatic Calibration System
The baseline, wavelength, dark current can be calibrated automatically to keep device in good running conditions.
◆ 6 inches Large LCD display
Equipped with a 6 inches large LCD display to show results and curves directly on the screen.
◆ Easy Data Output and Printing
Equipped with USB port to connect with PC, easy to export and print data and analysis report.
[Technical Specification]
Model | UV-9000 | UV-9000A | UV-9000S |
Optical System | Double Beam (1200 Lines/mm Grating) | ||
Wavelength Range | 190-1100nm | ||
Bandwidth | 1.8nm | 1.0nm | 0.5, 1.0, 2.0, 4.0 nm |
Wavelength Accuracy | ±0.1nm (D2 656.1nm), ±0.3nm @all | ||
Wavelength Repeatability | 0.2nm | ||
Photometric Accuracy | ±0.2%T(0-100%T) | ||
Photometric Repeatability | ≤0.1%T(0-100%T) | ||
Photometric Range | -0.3-3A, 0-200%T, 0-9999C | ||
Stability | ± 0.0003A/h @ 500nm | ||
Baseline Flatness | ± 0.001A/h | ||
Noise | ± 0.0005A/h | ||
Stray Light | ≤0.03%T @ 220nm, 360nm | ||
Data Output Port | USB | ||
Printer Port | Parallel Port | ||
Display | 320*240 Dots LCD | ||
Lamps | Tungsten Lamp & Deuterium Lamp | ||
Detector | Silicon Photodiode | ||
Power Requirements | AC 220V/50Hz or AC 110V/60Hz | ||
Dimension | 625*430*206mm | ||
Weight | 32kg/32kg/34kg | ||
[Standard Accessories]
Item | Description | Quantity | Unit |
1 | Spectrophotometer | 1 | set |
2 | 1cm Glass cuvette | 4 | pcs |
3 | Power cord | 1 | pcs |
4 | User’s Manual | 1 | pcs |
5 | 1 cm quartz cuvette | 2 | pcs |
6 | Dust Cover | 1 | pcs |
7 | Software | 1 | pcs |

Business Type:Manufacturer
Online Supplier